Baby ICU Incubator Medical Equipment Neonatal Infant Transport Incubator For Ambulance Use
Features
1 Air mode and Baby mode controlled by micro-computer;
2 Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
3 Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
4 Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it;
5 Natural air flow humidity;
6 The height of whole unit can be adjusted;
7 With Oxygen cylinder and Oxygen supply configuration;
8 Observation lamp.
Mode of control |
Air mode and baby mode controlled by micro-computer |
Control range of air temperature |
25°C-37°C(Override mode 37-38°C) |
Control range of baby temperature |
34°C-37°C(Override mode 37-37.5°C) |
Accuracy of skin temperature sensor |
≤0.3°C |
Variability of temperature |
≤1.0°C |
Uniformity of temperature |
≤1.5°C |
Noise inside hood |
≤60dB(A) |
Failure alarm |
Over-temperature alarm |
Temperature deviation alarm |
Sensor alarm |
Motor Fan Failed Alarm |
Power failure Alarm and so on |
Running time of internal battery |
90min (1 internal battery) |
AC Power supply |
AC220V-230V, 50Hz |
DC Power supply |
DC12V/10A or DC24V/6A |
Input power |
400VA |
Main Body |
Weight |
80kg |
Dimension |
109*65.5*76cm |
Cabinet |
Weight |
45kg |
Dimension |
138*61*40cm |