Customization: | Available |
---|---|
Certification: | ISO13485 |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Dimension | 146 x 193 x 95mm |
Infusion set | Applicable to GB 8368-2005 disposals; diametersΦ3.5-4.5mm, thickness 0.8-1.2mm |
Infusion mode | Rate mode, Time mode, Body weight mode, Sequential mode |
Flow rate range | 0.1~1500ml/h |
Accumulated volume | 0.1~9999 ml |
VTBI | 0.1~999.9 ml, increment is 0.1 ml; 1000-9999ml, increment is 1ml |
Increment | 0.1~999.9 ml/h, increment is 0.1 ml/h; 1000~1500 ml/h, increment is 1 ml/h |
Accuracy | ≤ ±5% |
KVO rate | 0.1-5.0 ml/h, adjustable, increment is 0.1ml/h |
Bolus rate | 0.10-1500ml/h |
Purge rate | 800ml/h |
Occlusion | Low 150±125 mmHg (20kPa); Medium 525±125 mmHg (70kPa); High 900±180 mmHg (120 kPa) |
Air bubble detection | Single bubble detection: 50, 100, 250, 500, 800l μl Accumulated bubble detection: 0.1-4.0ml/h, adjustable |
Screen | 3.0'' LCD monochrome screen, 240 x 128 pixels |
Titration | Yes, increment is 0.1ml/h |
Keypad lock | Yes |
Pause | Yes |
Self-test system | Yes |
Last therapy parameter memory | Yes |
Drug library | 200 drugs |
History record | 1500 events |
Weight | <1.6kg (not including clamp) |
Handle | Embedded handle |
Mounting type | Standard: Vertical Pole |
Alarm Volume | 1-8 levels selectable |
Language | 8, Chinese, English, Spanish, French, Russian, Turkish, Portuguese, German |
Alarm | VTBI done, VTBI near done, KVO finish, Bottle empty, door open, IV set not installed, occlusion, air in line, drop error, system error, system abnormal, batter y low, batter y empty, reminder, AC power disconnection, standby end |
Safety Requirement | |
AC power | 100-240V (50/60Hz, 0.8-0.28A) |
DC power | 10-16V (4.0-2.5A) |
Battery type | Lithium |
Battery operation time | >4 hours@25ml/h |
Charging time | ≤6 hours@100% |
Classification | CF, Class I |
IP Grade | IP 34 |
Temperature | 5~40ºC for operating, -20~60ºC for storage |
Humidity | 15~95% for operating, 10~95% for storage |
Air pressure | 57~106kPa for operating, 50~106kPa for storage |
Interface | |
Data interface | Supported, RS232 |
Nurse Call | Supported |
DC input | Supported |