Warmer In Nicu Resuscitaire Neonatal Infant Radiant Warmer
Brief Introduction:
Three control modes: Pre-warm mode control, manual mode control and baby mode control by micro-computer;
Set temperature and skin temperature can be displayed separately;
Main body is made of stainless steel;
Double sides phototherapy radiant box;
Acrylic glass bassinet;
APGAR timer;
Timer to count phototherapy time;
Failure alarm indication;
RS232 connector;
Disposable skin temperature sensor is optional.
Standard Configuration:
Warmer module, temperature controller, bassinet, phototherapy radiant box, and the stand.
Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Mode of control: Pre-warm mode, manual mode and baby mode are controlled by micro-computer
Baby mode temperature control range: 34.5-37.5ºC
Skin temperature display range: 5ºC-65ºC
Accuracy of skin temperature sensor: ±0.2ºC
Temperature uniformity of mattress: ≤2ºC
Dominant wavelength of blue light: 400-550nm
Timing range of blue light: 0~9999hrs 59min
APGAR timer: audible and visible alarm when the device runs to 50″~1′,4′50″~5′,9′50″~10′;
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.
Average value of general bilirubin radiation on the mattress effective range:
≥250uW/cm2(the above phototherapy equipment)
≥600uW/cm2(the nether phototherapy equipment)
Bilirubin uniformity of radiating on the mattress effective range: >0.4