Medical Equipment Pediatric Incubator Manufacturers Baby Neonatal Infant Incubator
Specifications of hood |
BI-3000 |
Mattress to hood |
450 mm |
Observation side of hood |
Two incline |
Operate port |
6 (2 Iris ports) |
Tubing grommets |
4 |
Front access panel |
826*310 mm |
Mattress Tray Size |
630*360 mm |
Technical Specification
Power requirement |
≤400VA |
Temperature control mode |
Air and skin servo control |
Air Temp Control Range |
20°C 37°C |
Temperature override mode to |
20°C 39°C |
Skin Temp Control Range |
30°C 37°C |
Temperature override mode to |
30°C 38°C |
Skin Temp sensor accuracy |
≤0.3°C |
Humidity display range |
5% 99% |
Temperature fluctuation |
≤0.5°C |
Mattress temperature uniformity |
≤0.8°C |
Temperature rise time |
≤30 mim |
Noise level within hood environment |
≤55db |
Alarm
Power failure alarm |
Yes |
Fan failure alarm |
Yes |
Sensor failure alarm |
Yes |
High air temperature alarm |
+1.5°C |
Low air temperature alarm |
-2.5°C |
High skin temperature alarm |
+1.0°C |
Low skin temperature alarm |
-1.0°C |
Over air temperature alarm |
≤38.0°C; ≤40.0°C |
Over skin temperature alarm |
≤38.0°C; ≤39.0°C |
System failure alarm |
Yes |
Alarm silence function |
Yes |
Operational condition
Environment temperature |
+20°C +30°C |
Environment relative humidity |
5% 99% |
Environment air velocity of flow |
<0.3m/s |
Others
RS-232 Connector |
Yes |
Rail |
Yes |
Infant bed tilt adjusting |
±10° step-less |
Mattress tray Max. load |
10 kg |
Transfusion pole Max. load |
2 kg |
Cabinet Max. load |
6 kg |