| Oxygen concentration | Adjusting range:21%-100% |
| Monitoring range:15%-100% | |
| CPAP | Adjusting range:0cmH2O-15cmH2O |
| Monitoring range:0cmH2O-60cmH2O | |
| Continuous flow | Adjusting range:2L/min-20L/min |
| Monitoring range:2L/min-20L/min | |
| Manual Ventilation | Output flow:20L/min±5L/min |
| Air compressor | Continuous output flow:≥25L/min Peak output flow:≥120L/min |
Graphical display: |
|
| P-T(pressure - time) | |
Size Wooden case packing size:L 670*W 700*H 1410mm , GW:90KG ; NW:57.2KG Alarm and protection |
|
| The AC power failure alarm | Power failure or noconnection |
| Internal battery backup low voltage alarm | <11.3±0.3V |
| High oxygen concentration alarm Low oxygen concentration alarm |
19%-100% 15%-99% |
| High Airway pressure alarm Low air input pressure alarm Low oxygen input pressure alarm Air and oxygen blender alarm |
1.5kPa-6kPa <200kPa <200kPa When any gas source lack |
| When differential pressure> 0.1mPa | |
| Compression pump alarm |
Excess temperature |
| Working conditions | |
| Gas source | O2,AIR |
| Pressure | 280kPa-600kPa |
| Voltage | 100-240V |
| Power frequency | 50Hz±1Hz |
| Input power | Main engine:≤25W Air compressor:≤675W |